BMC Gastroenterology (Jun 2010)
Colon Capsule Endoscopy compared to Conventional Colonoscopy under routine screening conditions
Abstract
Abstract Background Colonoscopy (CSPY) for colorectal cancer screening has several limitations. Colon Capsule Endoscopy (PillCam Colon, CCE) was compared to CSPY under routine screening conditions. Methods We performed a prospective, single-center pilot study at a University Hospital. Data were obtained from November 2007 until May 2008. Patients underwent CCE on Day 1 and CSPY on Day 2. Outcomes were evaluated regarding sensitivity and specificity of polyp detection rate, with a significance level set at >5 mm. Results 59 individuals were included in this study, the results were evaluable in 56 patients (males 34, females 22; median age 59). CCE was complete in 36 subjects. Polyp detection rate for significant polyps was 11% on CSPY and 27% on CCE. 6/56 (11%) patients had polyps on CSPY not detected on CCE (miss rate). Overall sensitivity was 79% (95% confidence interval [CI], 61 to 90), specificity was 54% (95% CI, 35 to 70), positive predictive value (PPV) was 63% and negative predictive value (NPV) was 71%. Adjusted to significance of findings, sensitivity was 50% (95% CI, 19 to 81), specificity was 76% (95% CI, 63 to 86), PPV was 20% and NPV was 93%. Conclusion In comparison to the gold standard, the sensitivity of CCE for detection of relevant polyps is low, however, the high NPV supports its role as a possible screening tool. Trial Registration NCT00991003.