Journal of Ardabil University of Medical Sciences (Oct 2023)
A Novel Design of a Non-Enzymatic Biosensor for the Detection of Hydrogen Peroxide Based on MWCNT/Co3O4/Hemoglobin Structures (Original Article)
Abstract
Background: Determining the concentration of hydrogen peroxide in liquids and biological samples is very important because of its effects on human health. This study aimed to design a new electrochemical biosensor based on hemoglobin to detect hydrogen peroxide in serum samples. Methods: In this study, a basic science, a biosensor based on modifying the glassy carbon electrode surface with a nanocomposite consisting of cobalt oxide nanoparticles and multi-walled carbon nanotube functionalized with a carboxyl group (MWCNT/Co3O4) and hemoglobin stabilized on this nanocomposite was made as a biological recognition element. Results: In optimal conditions, the biosensor was used to measure different concentrations of hydrogen peroxide. The designed biosensor showed a wide linear response range from 10 μM to 500 μM, a detection limit of 0.512 μM, and high reproducibility and stability. Conclusion: In this innovative research work, MWCNTs/Co3O4 nanocomposite was used to make a diagnostic biosensor. The presented biosensor showed an acceptable performance in the measurement of hydrogen peroxide in serum samples and laboratory solutions.