Journal of Dental Sciences (Dec 2011)

Pullout strengths of orthodontic palatal mini-implants tested in vitro

  • Ju-Hui Wu,
  • Huang-Chi Wang,
  • Chun-Ming Chen,
  • Pei-Chen Lu,
  • Sheng-Tsung Lai,
  • Kun-Tsung Lee,
  • Je-Kang Du

DOI
https://doi.org/10.1016/j.jds.2011.09.003
Journal volume & issue
Vol. 6, no. 4
pp. 200 – 204

Abstract

Read online

Background/purpose: New modified mini-implants are used in orthodontic practice to reinforce palatal anchorage. The aim of this study was to evaluate the anchorage strengths of palatal mini-implants in terms of their vertical and horizontal pullout strengths. Materials and methods: Thirty palatal mini-implants (2 mm in diameter) of three brands (Absoanchor, Bio-Ray, and Lomas) were manually driven into artificial bone (Sawbones) to a depth of 5 mm. Their vertical and horizontal pullout strengths were measured using a material testing machine. The Kruskal–Wallis test was used to assess differences among brands (P<0.05). Results: The pullout strengths of all the brands were significantly greater than routine orthodontic forces. The vertical pullout strength of the Absoanchor mini-implants was the lowest among the tested brands, and the horizontal pullout strengths of the Lomas and Absoanchor mini-implants were significantly higher than that of the Bio-Ray mini-implant. There was no significant relationship between the insertion torque and pullout strength in the vertical or horizontal directions. Conclusion: The pullout strengths of mini-implants were significantly greater than normal orthodontic forces. Moreover, no significant correlation was found between the insertion torque and pullout strength.