Technical Innovations & Patient Support in Radiation Oncology (Mar 2020)

Laser-free pelvic alignment in an online adaptive radiotherapy environment

  • Helen Barnes,
  • Jonathan Mohajer,
  • Alex Dunlop,
  • Gillian Adair Smith,
  • Trina Herbert,
  • Rebekah Lawes,
  • Alison Tree,
  • Helen McNair

Journal volume & issue
Vol. 13
pp. 21 – 23

Abstract

Read online

The MR-Linac (MRL) provides a novel treatment modality that enables online adaptive treatments, but also creates new challenges for patient positioning in a laser-free environment. The accuracy and duration of prostate patient set-up on the MRL using two different methods for patient alignment was determined to establish standard of practice on the MRL. Differences in set-up accuracy were significant in the longitudinal direction and are accounted for in online plan adaption. Both methods recorded similar set-up times. The vendor recommended alignment method involves less manipulation of the patient and will be adopted as the standard positioning method for prostate and other pelvic patients on the MRL in future. Keywords: Laser-free, ART, MR Linac, Alignment