Effect of Fluorescent Condensation Plates (CaWO4) on Radiation Dose Received at X-Ray.

  • Qusay Aldulayme,
  • Muhsin Albadrane

DOI
https://doi.org/10.33899/edusj.2009.57777
Journal volume & issue
Vol. 22, no. 3
pp. 105 – 114

Abstract

Read online

Abstract The present study aims to investigate the effect of using (CaWO4) as intensifying screen on radiation dose received by patients exposed to dental investigation by x-ray. It was found that intensifying screen increases, the radiation dose from )4-6( times according to the properties of the intensifying materials used. The optical density of the x-ray film was measured by using densitometer. According to our measurements, a decrease in radiation dose was approximately 24% for the same condition .Am241 was used because its energy (59.5keV) is close to the energy of x-ray in dental radiography.