IEEE Access (Jan 2022)

Metamaterial Liner for MRI Excitation—Part 2: Design and Performance at 4.7T

  • Adam Mitchell Maunder,
  • Ashwin K. Iyer,
  • Nicola De Zanche

DOI
https://doi.org/10.1109/ACCESS.2022.3167764
Journal volume & issue
Vol. 10
pp. 41678 – 41692

Abstract

Read online

The theoretical foundations of a metamaterial (MM) liner for the MRI bore that facilitates the propagation of reduced-cutoff cylindrical waveguide modes are presented in the Part 1 companion paper to this work. Here, in Part 2, the practical design and modelling of the novel MM liner is applied to a body MRI radio-frequency (RF) transmitter for 4.7T. An equivalent network mesh model is developed to reduce the distributed structure of the MM to one that uses lumped discrete tuning elements. The close match between full-wave simulation and the network model is demonstrated by comparison of the longitudinal propagation dispersion curves and input impedance. The application of the methods developed for analysis of the MM liner behavior are demonstrated in the design of a practical MM liner as an effective MRI RF transmitter. The liner’s simulated performance was evaluated using three key metrics of MR radio-frequency (RF) coil performance: transmit efficiency, the variation of the transmit field (a measure of homogeneity), and the 10g averaged local specific absorption rate (SAR). These metrics are compared to those of the commonly used birdcage (BC) coil. The main advantage of the MM liner is the 29.6% reduction in maximum SAR normalized to transmit field relative to the BC coil, with comparable transmit efficiency and homogeneity. Thus, the liner can potentially replace the BC coil as an RF transmitter and provide an enhanced safety margin, or allow the use of faster, more SAR-aggressive imaging sequences.

Keywords