Wellcome Open Research (Apr 2018)

Evaluation of the Luminex xTAG Respiratory Viral Panel FAST v2 assay for detection of multiple respiratory viral pathogens in nasal and throat swabs in Vietnam [version 2; referees: 2 approved]

  • Vu Thi Ty Hang,
  • Nguyen Thi Han Ny,
  • Tran My Phuc,
  • Pham Thi Thanh Tam,
  • Dang Thao Huong,
  • Ho Dang Trung Nghia,
  • Nguyen Tran Anh Vu,
  • Pham Thi Hong Phuong,
  • Nguyen Van Xang,
  • Nguyen Dong,
  • Pham Nhu Hiep,
  • Nguyen Van Hung,
  • Tran Tinh Hien,
  • Maia Rabaa,
  • Guy E. Thwaites,
  • Stephen Baker,
  • Le Van Tan,
  • H.Rogier van Doorn,
  • VIZIONS consortium

DOI
https://doi.org/10.12688/wellcomeopenres.12429.2
Journal volume & issue
Vol. 2

Abstract

Read online

Background: Acute respiratory infections (ARI) are among the leading causes of hospitalization in children ≤5 years old. Rapid diagnostics of viral pathogens is essential to avoid unnecessary antibiotic treatment, thereby slowing down antibiotic-resistance. We evaluated the diagnostic performance of the Luminex xTAG Respiratory Viral Panel FAST v2 against viral specific PCR as reference assays for ARI in Vietnam. Methods: Four hundred and forty two nose and throat swabs were collected in viral transport medium, and were tested with Luminex xTAG Respiratory Viral Panel FAST v2. Multiplex RT-PCR and single RT-PCR were used as references. Results: Overall, sensitivity of the Luminex against reference assays was 91.8%, 95% CI 88.1-94.7 (270/294), whilst 112/6336 (1.8%, 95% CI, 1.4-2.1) of pathogens were detected by the Luminex, but not by reference assays. Frequency of pathogens detected by Luminex and reference assays was 379 and 292, respectively. The diagnostic yield was 66.7% (295/442, 95%CI 62.1-71.1%) for the Luminex assay and 54.1% (239/442, 95% CI, 49.3-58.8%) for reference assays. The Luminex kit had higher yields for all viruses except influenza B virus, respiratory syncytial virus, and human bocavirus. High agreements between both methods [mean (range): 0.91 (0.83-1.00)] were found for 10/15 viral agents. Conclusions: The Luminex assay is a high throughput multiplex platform for rapid detection of common viral pathogens causing ARI. Although the current high cost may prevent Luminex assays from being widely used, especially in limited resource settings where ARI are felt most, its introduction in clinical diagnostics may help reduce unnecessary use of antibiotic prescription.

Keywords