Archives of Clinical and Experimental Surgery (Jun 2019)

A polyethylene liner dissociation case report in depuy pinnacle cup. An impingement problem?

  • Alvaro Martinez-Ayora,
  • Guillermo Sosa-Gonzalez,
  • Carlos Luis Vidal-Fernandez

DOI
https://doi.org/10.5455/aces.20190318055641
Journal volume & issue
Vol. 8, no. 2
pp. 49 – 52

Abstract

Read online

The modular components facilitate joint replacement surgery but are associated with potential complications not present in the previous components. We present a case of polyethylene dissociation due to rupture of the anti-rotational blocks in a total hip prosthesis two years after surgery. We think about the possible causes and risk factors. Despite its low incidence, the dissociation between polyethylene and the acetabular cup is a catastrophic complication. The femoroacetabular impingement seems to play a key role. [Arch Clin Exp Surg 2019; 8(2.000): 49-52]