Scientific Reports (Jul 2021)
Intraoral scanners for capturing the palate and its relation to the dentition
Abstract
Abstract Proper superimposition of intraoral scan generated 3D models enables detailed assessment of soft and hard tissue surface changes. This requires accurate 3D models and stable structures as superimposition references. In the maxilla, different reference areas have been proposed, mostly located at the palatal region. In this in vivo study we evaluated the precision of two intraoral scanners (TRIOS 3, 3Shape and CS 3600, Carestream) at the maxilla, focusing on the palate itself and also on its spatial relation to the dentition, following palatal superimposition. Precision was tested through the superimposition of repeated scans on the palate and the dental arch. Overall, the median precision of both scanners was high (< 0.1 mm). Scanner precision was comparable when the palatal area was tested individually. However, TRIOS 3 showed higher precision regarding the assessment of the dental arch, following superimposition of repeated models on the palate (median difference: approximately 40 μm). In few cases, local areas of higher imprecision were present for both scanners, exceeding 0.3 mm. Thus, scanner precision seems to be high in small, but slightly reduced considering larger areas, with differences between scanners. However, the effect on individual tooth position relative to the palate was for both scanners limited.