Medical Devices: Evidence and Research (Jan 2015)

Device-length changes and implant function following surgical implantation of the KineSpring in cadaver knees

  • McNicholas MJ,
  • Gabriel SM,
  • Clifford AG,
  • Hasler EM

Journal volume & issue
Vol. 2015, no. default
pp. 47 – 56

Abstract

Read online

Michael J McNicholas,1 Stefan M Gabriel,2 Anton G Clifford,2 Evelyne M Hasler2 1Aintree University Hospital, Teaching Hospital, Major Trauma Centre, NHS Foundation Trust, Liverpool, UK; 2Moximed, Hayward, CA, USA Introduction: The KineSpring implant system has been shown to provide load reductions at the medial compartment of the knee, and has demonstrated clinical success in reducing pain and increasing function in patients with medial knee osteoarthritis. These results depend on the ability of the KineSpring to rotate, lengthen, and shorten to accommodate knee motions, and in response to knee position and loading. Purpose: The present study was undertaken to determine length changes of the implanted KineSpring in response to a range of knee positions, external knee loads, and placements by different orthopedic surgeons. Materials and methods: KineSpring system components were implanted in ten cadaver leg specimens by ten orthopedic surgeons, and absorber-length changes were measured under combined loading and in different positions of the knee. Results and conclusion: Spring compression consistent with knee-load reduction, and device lengthening and shortening to accommodate knee loads and motions were seen. These confirm the functionality of the KineSpring when implanted medially to the knee. Keywords: KineSpring, knee, function, preservation, offloading, osteoarthritis