Indian Journal of Ophthalmology (Jan 2023)

An evaluation of the accuracy of toric intraocular lens power calculation based on measured total corneal refractive power

  • Zhi Yang,
  • Yanwen Zhou,
  • Tianru Jin,
  • Jingjing Li

DOI
https://doi.org/10.4103/ijo.IJO_1539_22
Journal volume & issue
Vol. 71, no. 2
pp. 541 – 546

Abstract

Read online

Purpose: To evaluate a method using measured values of total corneal refractive power (TCRP) for a manufacturer's online calculator by comparing it with the Barrett toric calculator (BTC) and Kane toric calculator (KTC) combined with simulated keratometry values (SimK). Methods: This was a retrospective case series. Patient records were reviewed to identify the patients who had biometry with the IOL Master 700 and Pentacam recorded before toric IOL implantation and refractive follow-up data after implantation. The predicted error in residual astigmatism was calculated by vector analysis according to the calculation methods and the measurements used. Results: A total of 70 eyes of 56 patients were included. The mean absolute astigmatism prediction errors were 0.6 ± 0.32, 0.59 ± 0.35, and 0.61 ± 0.35 D for the ATCTCRP, BTCSimK, and KTCSimK calculators, respectively (P = 0.934), and the centroid of the prediction errors were 0.3 D @ 178°, 0.11 D @ 102°, and 0.09 D @ 147°, respectively (P = 0.23). In the with-the-rule subgroup, the centroid of the prediction error was 0.34 D @ 176° for ATCTCRP and was the highest among the three calculation methods (P = 0.046). Conclusion: The ATCTCRP, BTCSimK, and KTCSimK calculators had similar performance with regards to their astigmatism prediction accuracy. The ATCTCRP calculator combined with 4.0-mm apex/ring readings of TCRP was slightly intended to result in against-the-rule residual astigmatism.

Keywords