Diagnostics (Jul 2022)

Using the Axial Oblique View of Computed Tomography (CT) in Evaluating Femoral Anteversion: A Comparative Cadaveric Study

  • Kwang-Soon Song,
  • Chang-Jin Yon,
  • Yu-Ran Heo,
  • Jae-Ho Lee,
  • Seung-Bo Lee,
  • Yeon-Kyoung Ko,
  • Kyung-Jae Lee,
  • Si-Wook Lee

DOI
https://doi.org/10.3390/diagnostics12081820
Journal volume & issue
Vol. 12, no. 8
p. 1820

Abstract

Read online

Twenty-five cadaveric adult femora’s anteversion angles were measured to develop a highly efficient and reproducible femoral anteversion measurement method using computed tomography (CT). Digital photography captured the proximal femur’s two reference lines, head-to-neck (H-N) and head-to-greater trochanter (H-G). Six reference lines (A/B in transverse section; C, axial oblique section; D/E, conventional 3D reconstruction; and M, volumetric 3D reconstruction) from CT scans were used. The posterior condylar line was used as a distal femoral reference. As measured with the H-N and H-G lines, the anteversion means were 10.43° and 19.50°, respectively. Gross anteversion measured with the H-G line had less interobserver bias (ICC; H-N = 0.956, H-G = 0.982). The 2D transverse and volumetric 3D CT sections’ B/M lines were consistent with the H-N line (p: B = 0.925, M = 0.122) and the 2D axial oblique section’s C line was consistent with the H-G line (p p < 0.05). Among several CT scan femoral anteversion measurement methods, the novel anteversion angle measurement method using CT scans’ axial oblique section was approximated with actual gross femoral anteversion angle from the femoral head to the greater trochanter.