BMC Medical Imaging (Feb 2023)

A quantitative clinical evaluation of simultaneous reconstruction of attenuation and activity in time-of-flight PET

  • Haiqiong Zhang,
  • Jingnan Wang,
  • Nan Li,
  • Yue Zhang,
  • Jie Cui,
  • Li Huo,
  • Hui Zhang

DOI
https://doi.org/10.1186/s12880-023-00987-7
Journal volume & issue
Vol. 23, no. 1
pp. 1 – 13

Abstract

Read online

Abstract Background The maximum likelihood activity and attenuation (MLAA) reconstruction algorithm has been proposed to jointly estimate tracer activity and attenuation at the same time, and proven to be a promising solution to the CT attenuation correction (CT-AC) artifacts in PET images. This study aimed to perform a quantitative evaluation and clinical validation of the MLAA method. Methods A uniform cylinder phantom filled with 18F-FDG solution was scanned to optimize the reconstruction parameters for the implemented MLAA algorithm. 67 patients who underwent whole-body 18F-FDG PET/CT scan were retrospectively recruited. PET images were reconstructed using MLAA and clinical standard OSEM algorithm with CT-AC (CT-OSEM). The mean and maximum standardized uptake values (SUVmean and SUVmax) in regions of interest (ROIs) of organs, high uptake lesions and areas affected by metal implants and respiration motion artifacts were quantitatively analyzed. Results In quantitative analysis, SUVs in patient’s organ ROIs between two methods showed R 2 ranging from 0.91 to 0.98 and k ranging from 0.90 to 1.06, and the average SUVmax and SUVmean differences between two methods were within 10% range, except for the lung ROI, which was 10.5% and 16.73% respectively. The average SUVmax and SUVmean differences of a total of 117 high uptake lesions were 7.25% and 7.10% respectively. 20 patients were identified to have apparent respiration motion artifacts in the liver in CT-OSEM images, and the SUVs differences between two methods measured at dome of the liver were significantly larger than measured at middle part of the liver. 10 regions with obvious metal artifacts were identified in CT-OSEM images and the average SUVmean and SUVmax differences in metal implants affected regions were reported to be 52.90% and 56.20% respectively. Conclusions PET images reconstructed using MLAA are clinically acceptable in terms of image quality as well as quantification and it is a useful tool in clinical practice, especially when CT-AC may cause respiration motion and metal artifacts. Moreover, this study also provides technical reference and data support for the future iteration and development of PET reconstruction technology of SUV accurate quantification.

Keywords