Electronics (May 2019)

Analysis and Design of an Energy Verification System for SC200 Proton Therapy Facility

  • Pengyu Wang,
  • Jinxing Zheng,
  • Yuntao Song,
  • Wuquan Zhang,
  • Ming Wang

DOI
https://doi.org/10.3390/electronics8050541
Journal volume & issue
Vol. 8, no. 5
p. 541

Abstract

Read online

The purpose of this study is to provide an energy verification method for the nozzle of the SC200 proton therapy facility to ensure safe redundancy of treatment. This paper first introduces the composition of the energy selection system of the SC200 proton therapy facility. Secondly, according to IEC60601 standard, the energy verification requirement that correspond to 1 mm error in water is presented. The allowable difference between the measured magnetic field and the reference are calculated based on the energy verification requirements to select the field resolution of the Hall probe. To ensure accuracy and stability, two Hall probes are mounted on the dipole to monitor the magnetic field strength to verify the proton beam energy in real time. In addition, the test results of the residual field of the dipole show that the probe system meets the accuracy requirements of energy verification. Furthermore, the maximum width of the slit of the energy selection system in accordance with the IEC standard at the corresponding energy is calculated and compared with the actual position of the movable slit to verify the momentum divergence of the proton beam. Finally, we present an energy verification method.

Keywords