Scientific Reports (Jan 2024)

Combination of high resolution MRI with 3D-printed needle guides for ex vivo myocardial biopsies

  • Simon Reiss,
  • Julien Thielmann,
  • Johannes Fischer,
  • Thomas Lottner,
  • Alexander Maier,
  • Dirk Westermann,
  • Constantin von zur Mühlen,
  • Timo Heidt,
  • Michael Bock

DOI
https://doi.org/10.1038/s41598-023-50943-2
Journal volume & issue
Vol. 14, no. 1
pp. 1 – 9

Abstract

Read online

Abstract Magnetic resonance imaging (MRI) provides a multitude of techniques to detect and characterize myocardial infarction. To correlate MRI findings with histology, in most cases terminal animal studies are performed; however, precise extraction and spatial correlation of myocardial tissue samples to MRI image data is difficult. In this proof of concept study, we present a 3D-printing technique to facilitate the extraction of tissue samples from myocardial regions. Initially, seven pig hearts embedded in formaldehyde were imaged on a clinical 3 T system to define biopsy targets on high resolution ex vivo images. Magnitude images and R2*-maps acquired with a 3D multi-echo gradient echo sequence and 0.58 mm isotropic resolution were used to create digital models of the cardiac anatomy. Biopsy guides were 3D-printed to steer the extraction of myocardial samples. In total, 61 tissue samples were extracted with an average offset of the tissue sample location from the target location of 0.59 ± 0.36 mm. This offset was not dependent on the distance of the target point to the epicardial surface. Myocardial tissue could be extracted from all samples. The presented method enables extraction of myocardial tissue samples that are selected by ex vivo MRI with submillimeter precision.