Indian Journal of Anaesthesia (Jan 2020)

Comparison of effectiveness of the piston-pump method versus the pressure-infusor method for rapid infusion of crystalloids: A bench study

  • Wataru Hashimoto,
  • Ichiro Takenaka,
  • Keisuke Yasunami,
  • Tomoko Minami,
  • Haruhiko Sano

DOI
https://doi.org/10.4103/ija.IJA_864_20
Journal volume & issue
Vol. 64, no. 12
pp. 1059 – 1063

Abstract

Read online

Background and Aims: The piston-pump method is a simple method for rapid administration of fluids but some problems are unsolved. We compared the effectiveness of using the piston-pump method with that of the pressure-infusor method. Methods: Twelve anaesthetists were classified randomly into the piston-pump and pressure-infusor groups. They were asked to infuse 500 ml of saline three times successively through a 16-G intravenous cannula as rapidly as possible using a pump with a 50-ml syringe or a pressure-infusor at 300 mmHg. The time taken for infusion and the maximum or minimum pressure in the infusion circuit and substitute vessel were measured. Bacterial culture of the saline infused sterilely was performed to estimate bacterial contamination. Results: The pressure-infusor group led to faster infusion of 500 ml of saline (233 ± 19 s) than the piston-pump group (301 ± 48 s) (P 200 (P < 0.01) and in the substitute vessel was 5 ± 1 and 17 ± 7 (P < 0.01) in the pressure-infusor and piston-pump groups, respectively. A pressure of <-200 mmHg occurred at all infusion attempts in the piston-pump group. Bacterial contamination was not observed in either group. Conclusion: If fluids must be administered rapidly, the pressure-infusor method is more efficient than the piston-pump method because the latter is less effective in infusing fluids rapidly and associated with excessive positive and negative pressure in the infusion circuit.

Keywords