Biomimetics (Aug 2023)

Different Methods for Assessing Tooth Colour—In Vitro Study

  • Susana Dias,
  • Joana Dias,
  • Ruben Pereira,
  • João Silveira,
  • António Mata,
  • Duarte Marques

DOI
https://doi.org/10.3390/biomimetics8050384
Journal volume & issue
Vol. 8, no. 5
p. 384

Abstract

Read online

Colour assessment using digital methods can yield varying results, and it is important for clinicians to recognize the potential variability intra and inter-device. This study aimed to compare the L*a*b* values of VITA Classical (VC) and VITA Toothguide 3D-MASTER (VM) guides using two methods, SpectroShade (SS) and eLAB. Thirty-four measurements per tab were performed by a single operator across three batches of each guide. Intraclass correlation coefficients (ICC) between batches were calculated. Values 0.90 were classified as poor, moderate, good, and excellent reliability, respectively. Results were reported as mean and standard deviation of the L*a*b* values and respective colour differences (ΔE00) for each tab and method. Statistical analyses were performed with an independent t-test, α = 0.05. ICC values between batches were excellent for all L*a*b*, except for a* component in eLAB. There were statistically significant differences between methods in most L*a*b* values. The intra-device mean ΔE00 was 0.5 ± 0.6 for VC, 0.5 ± 0.8 for VM in SS, 1.1 ± 0.8 for VC, 1.1 ± 0.9 for VM in eLAB. The mean ΔE00 inter-device was 4.9 ± 1.7 for VC, 5.0 ± 1.7 for VM. Both methods demonstrated good internal consistency, with high ICC values and low intra-device colour differences, but exhibited high variability between methods, higher for a* the component.