EPJ Web of Conferences (Jan 2022)

Alignment autocollimator-based microscope adjustment and its quality assessment

  • Krüger Jan,
  • Bergmann Detlef,
  • Sturm Matthias,
  • Häßler-Grohne Wolfgang,
  • Köning Rainer,
  • Bodermann Bernd

DOI
https://doi.org/10.1051/epjconf/202226610009
Journal volume & issue
Vol. 266
p. 10009

Abstract

Read online

We report a custom microscope setup whose mechanical and optical components are adjusted by the means of an alignment autocollimator (AAC). Residual centring and angular misalignments of the components towards the microscope’s optical axis are below 500 μm and 1 mrad, respectively. We further perform measurements of dot structures with diameters close to the diffraction limit (nominal diameter = 200 nm; chrome on glass mask) as suitable measures for the evaluation of the microscope’s adjustment and to determine/ visualize the optical aberrations, which affect the image formation of microscopes.