Investigation The Effect of Clearance and Body Weight on The Contact Pressure of Metal on PCU Hip Prosthesis using Finite Element Method
Abstract
A common problem with artificial hip replacements is increased wear of the material in contact. Materials that are in contact result in contact pressure caused by the patient's daily activities so that it triggers wear. This study adopts a finite element method (FEM) to predict wear of the artificial hip joint, by studying the behavior of a hip joint prosthesis that has clearance under a certain load. The aim of this study was to observe contact as a function of clearance and body weight. The modeling uses metal as femoral head and polycarbonate urethane (PCU) material as the acetabular cup. Contact modeling as a hard material in contact with a deformable material. Four variations of clearance (0.001, 0.005, 0.01, 0.016) and three variations of body weight (500N, 700N, and 1000N) were used in this study. The simulation results show that the lower the distance and weight, the lower the contact pressure.