Micromachines (Jun 2022)

Fluorescence Imaging Characterization of the Separation Process in a Monolithic Microfluidic Free-Flow Electrophoresis Device Fabricated Using Low-Temperature Co-Fired Ceramics

  • Pedro Couceiro,
  • Julián Alonso-Chamarro

DOI
https://doi.org/10.3390/mi13071023
Journal volume & issue
Vol. 13, no. 7
p. 1023

Abstract

Read online

A monolithic microfluidic free-flow electrophoresis device, fabricated using low-temperature co-fired ceramic technology, is presented. The device integrates gold electrodes and a 20 µm thick transparent ceramic optical window, suitable for fluorescence imaging, into a multilevel microfluidic chamber design. The microfluidic chamber consists of a 60 µm deep separation chamber and two, 50 µm deep electrode chambers separated by 10 µm deep side channel arrays. Fluorescence imaging was used for in-chip, spatial-temporal characterization of local pH variations in separation conditions as well as to characterize the separation process. The device allowed baseline resolution separation of a sample mixture of Fluorescein, Rhodamine 6G, and 4-Methylumbelliferone at pH 7.0, in only 6 s, using 378 V.s/cm. The results demonstrate the possibility of studying a chemical process using fluorescence imaging within the traditional fields of low-temperature co-fired ceramics technology, such as high-electrical-field applications, while using a simple fabrication procedure suitable for low-cost mass production.

Keywords