Results in Chemistry (Jan 2024)

Determination of four residual solvents in cefepime by headspace gas chromatography

  • Le Wang

Journal volume & issue
Vol. 7
p. 101269

Abstract

Read online

Objective: To establish a headspace gas chromatography method to determine the residual solvents of methanol, acetone, dichloromethane, and toluene in cefepime. Methods: Headspace gas chromatography using a DB-624 capillary column and temperature program was performed to detect four residual solvents in cefepime. The temperature program consisted of an initial temperature of 60 ℃maintained for 4 min, a temperature increased to 120 ℃ at 20 ℃·min−1 maintained for 3 min, and then a temperature increased to 200 ℃ at 40 ℃·min−1 maintained for 3 min. Results: Linear relationships were found between the peak areas and mass concentrations of the four solvents mentioned above within the defined ranges. The limits of detection for methanol, acetone, dichloromethane and toluene were 0.50, 0.15, 0.10 and 0.08 mg·L-1, respectively. The recoveries obtained using the standard addition method ranged from 98.41 % to 100.94 %. Conclusion: This method demonstrated high sensitivity and good reproducibility, and it can be used to effectively control the residual amounts of methanol, acetone, dichloromethane, and toluene, to further improve the quality of cefepime.