For this invited manuscript, an in-plane single-quartz-enhanced dual spectroscopy (IP-SQEDS)-based trace gas sensor was demonstrated for the first time. A single quartz tuning fork (QTF) was employed to combine in-plane quartz-enhanced photoacoustic spectroscopy (IP-QEPAS) with light-induced thermoelastic spectroscopy (LITES) techniques. Water vapor (H2O) was chosen as the target gas. Compared to traditional QEPAS, IP-SQEDS not only allowed for simple structures, but also obtained nearly three times signal amplitude enhancement.