The Mechanical Behavior of a Screwless Morse Taper Implant–Abutment Connection: An In Vitro Study

  • Aimen Bagegni,
  • Vincent Weihrauch,
  • Kirstin Vach,
  • Ralf Kohal

DOI
https://doi.org/10.3390/ma15093381
Journal volume & issue
Vol. 15, no. 9
p. 3381

Abstract

Read online

The use of screwless Morse taper implant–abutment connections (IAC) might facilitate the clinician’s work by eliminating the mechanical complications associated with the retention screw. The aim of this study is to evaluate the effect of artificial chewing on the long-term stability of screwless Morse taper IACs. Thirty-two implant abutments restored with an upper central incisor zirconia crown were used and divided into four groups according to the implant–abutment assembling manner (C1,H: screw retained (20 Ncm); C2: tapped; or C3: torqued (20 Ncm; the screws were removed before the dynamic loading)). All specimens were subjected to a cyclic loading (98 N) for 10 million chewing cycles. The survived samples were exposed to a pull-off force until failure/disassembling of the connection. All the samples revealed a 100% survival. Regarding the pull-off test, the screw-retained internal hexagonal IAC revealed significantly higher resistance to failure/disassembling (769.6 N) than screwless conical IACs (171.6 N–246 N) (p p > 0.05). The screw-retained hexagonal IAC showed the highest retention stability. The screw preload/retention in the conical IAC was lost over time in the group where the screws were kept in place during loading. Nevertheless, the screwless Morse taper IACs were stable for an extended service time and might represent a valid form of treatment for single-tooth replacement.