Journal of Lipid Research (Feb 1979)

Enzyme-linked immunoassay of ursodeoxycholic acid in serum.

  • S Ozaki,
  • A Tashiro,
  • I Makino,
  • S Nakagawa,
  • I Yoshizawa

Journal volume & issue
Vol. 20, no. 2
pp. 240 – 245

Abstract

Read online

A sensitive and specific enzyme-linked immunoassay for the measurement of ursodeoxycholic acid in human serum was developed. Ursodeoxycholic acid conjugated to alkaline phosphatase (from calf intestine) was used as a tracer. An antiserum to ursodeoxycholic acid serum was raised in rabbits using ursodeoxycholic acid–bovine serum albumin conjugate as an antigen. The binding required 1 hr at 42 degrees C; separation of the bound tracer was achieved by addition of a second antibody, and alkaline phosphatase activity of this bound complex was measured colorimetrically. The ratio of bound to total enzyme activity decreased linearly with a logarithmic increase in ursodeoxycholic acid concentration from 20 to 900 pmol. The specificity and sensitivity of this enzyme-linked immunoassay were similar to those of a radioimmunoassay reported previously. The serum ursodeoxycholic acid levels measured by this method correlated well with those determined by gas–liquid chromatography and radioimmunoassay. Based on these findings, this enzyme-linked immunoassay of bile acid might be useful as a tool for the routine clinical analysis of serum bile acids.