PLoS ONE (Jan 2016)

Bundled-Optode Method in Functional Near-Infrared Spectroscopy.

  • Hoang-Dung Nguyen,
  • Keum-Shik Hong,
  • Yong-Il Shin

DOI
https://doi.org/10.1371/journal.pone.0165146
Journal volume & issue
Vol. 11, no. 10
p. e0165146

Abstract

Read online

In this paper, a theory for detection of the absolute concentrations of oxy-hemoglobin (HbO) and deoxy-hemoglobin (HbR) from hemodynamic responses using a bundled-optode configuration in functional near-infrared spectroscopy (fNIRS) is proposed. The proposed method is then applied to the identification of two fingers (i.e., little and thumb) during their flexion and extension. This experiment involves a continuous-wave-type dual-wavelength (760 and 830 nm) fNIRS and five healthy male subjects. The active brain locations of two finger movements are identified based on the analysis of the t- and p-values of the averaged HbOs, which are quite distinctive. Our experimental results, furthermore, revealed that the hemodynamic responses of two-finger movements are different: The mean, peak, and time-to-peak of little finger movements are higher than those of thumb movements. It is noteworthy that the developed method can be extended to 3-dimensional fNIRS imaging.