Journal of Clinical Medicine (Oct 2022)

Accuracy of Six Intraocular Lens Power Calculations in Eyes with Axial Lengths Greater than 28.0 mm

  • Majid Moshirfar,
  • Kathryn M. Durnford,
  • Jenna L. Jensen,
  • Daniel P. Beesley,
  • Telyn S. Peterson,
  • Ines M. Darquea,
  • Yasmyne C. Ronquillo,
  • Phillip C. Hoopes

DOI
https://doi.org/10.3390/jcm11195947
Journal volume & issue
Vol. 11, no. 19
p. 5947

Abstract

Read online

The purpose of this study was to compare the accuracy of several intraocular (IOL) lens power calculation formulas in long eyes. This was a single-site retrospective consecutive case series that reviewed patients with axial lengths (AL) > 28.0 mm who underwent phacoemulsification. The Wang–Koch (WK) adjustment and Cooke-modified axial length (CMAL) adjustment were applied to Holladay 1 and SRK/T. The median absolute error (MedAE) and the percentage of eyes with prediction errors ±0.25 diopters (D), ±0.50 D, ±0.75 D, and ±1.00 D were used to analyze the formula’s accuracy. This study comprised a total of 35 eyes from 25 patients. The Kane formula had the lowest MedAE of all the formulas, but all were comparable except Holladay 1, which had a significantly lower prediction accuracy with either AL adjustment. The SRK/T formula with the CMAL adjustment had the highest accuracy in predicting the formula outcome within ±0.50 D. The newer formulas (BU-II, EVO, Hill-RBF version 3.0, and Kane) were all equally predictable in long eyes. The SRK/T formula with the CMAL adjustment was comparable to these newer formulas with better outcomes than the WK adjustment. The Holladay 1 with either AL adjustment had the lowest predictive accuracy.