BMC Ophthalmology (Nov 2019)

Comparability of three intraocular pressure measurement: iCare pro rebound, non-contact and Goldmann applanation tonometry in different IOP group

  • Min Chen,
  • Lina Zhang,
  • Jia Xu,
  • Xinyi Chen,
  • Yuxiang Gu,
  • Yuping Ren,
  • Kaijun Wang

DOI
https://doi.org/10.1186/s12886-019-1236-5
Journal volume & issue
Vol. 19, no. 1
pp. 1 – 10

Abstract

Read online

Abstract Background Measurement of intraocular pressure (IOP) is essential for glaucoma patients. Many factors such as central corneal thickness (CCT) can affect the accuracy of IOP measurement. The purpose of this study was to evaluate the agreement of IOP measured by non-contact tonometer (NCT), iCare pro rebound tonometer (iCare), and Goldmann applanation tonometer (GAT) in different IOP group. Methods This was a Hospital-based cross-sectional study. Two hundred subjects were enrolled in this study. All subjects underwent IOP measurement using an NCT–iCare–GAT sequence. Bland-Altman, Pearson correlation and intraclass correlation analysis were performed using SPSS 17.0 software. The influence of CCT on each IOP measurement methods was evaluated by linear regression analysis. Results The mean difference (Δ) of NCT–GAT did not differ from (Δ) iCare–GAT in IOP 30 mmHg group (P iCare pro > GAT.