npj Digital Medicine (Oct 2020)
Neuromuscular block monitoring by smartphone application (i-TOF© system): an observational pilot study
Abstract
Abstract Neuromuscular block monitoring is recommended by international guidelines to improve myorelaxation during surgery and reduce the risk of postoperative residual curarization. We conducted a pilot study to verify the efficacy of i-TOF, a wireless neuromuscular monitoring device connectable to a smartphone, comparing it with TOF WATCH SX. We enrolled 53 patients who underwent general anesthesia. For each patient, we recorded by both devices, in different time intervals, train-of-four (TOF) count/ratio after induction to general anesthesia (TI0–TI3) and during recovery (TR0–TR3). Moreover, post-tetanic count (PTC) was evaluated during deep neuromuscular block (TP0–TP2). We noticed no significant differences between the devices in recorded mean values of TOF ratio, TOF count, and PTC analyzed at time intervals for every phase of general anesthesia, although the i-TOF tends to an underestimation compared to TOF WATCH SX. For each patient, data sessions were successfully recorded by a smartphone. This aspect could be relevant for clinicians in order to have a stored proof of good clinical practice to be added on anesthesiologist records. By our results, i-TOF demonstrates a comparable efficacy to TOF WATCH SX, suggesting that it could be a proven alternative to standard devices for neuromuscular block monitoring. Further studies are needed to confirm our findings.