Radiation Oncology (Sep 2011)

Commissioning and early experience with a new-generation low-energy linear accelerator with advanced delivery and imaging functionalities

  • Fogliata Antonella,
  • Vanetti Eugenio,
  • Nicolini Giorgia,
  • Clivio Alessandro,
  • Cozzi Luca

DOI
https://doi.org/10.1186/1748-717X-6-129
Journal volume & issue
Vol. 6, no. 1
p. 129

Abstract

Read online

Abstract Background A new-generation low-energy linear accelerator (UNIQUE) was introduced in the clinical arena during 2009 by Varian Medical Systems. The world's first UNIQUE was installed at Oncology Institute of Southern Switzerland and put into clinical operation in June 2010. The aim of the present contribution was to report experience about its commissioning and first year results from clinical operation. Methods Commissioning data, beam characteristics and the modeling into the treatment planning system were summarized. Imaging system of UNIQUE included a 2D-2D matching capability and tests were performed to identify system repositioning capability. Finally, since the system is capable of delivering volumetric modulated arc therapy with RapidArc, a summary of the tests performed for such modality to assess its performance in preclinical settings and during clinical usage was included. Results Isocenter virtual diameter was measured as less than 0.2 mm. Observed accuracy of isocenter determination and repositioning for 2D-2D matching procedures in image guidance was Conclusions The results of the commissioning tests and of the first period of clinical operation, resulted meeting specifications and having good margins respect to tolerances. UNIQUE was put into operation for all delivery techniques; in particular, as shown by the pre-treatment quality assurance results, it enabled accurate and safe delivery of RapidArc plans.

Keywords