International Journal of Cardiology: Heart & Vasculature (Sep 2015)

Changes in the reference lumen size of target lesions before and after coronary stent implantation: Evaluation with frequency domain optical coherence tomography

  • Muneo Kurokawa,
  • Shiro Uemura,
  • Makoto Watanabe,
  • Yoko Dote,
  • Yu Sugawara,
  • Yutaka Goryo,
  • Tomoya Ueda,
  • Satoshi Okayama,
  • Michinori Kayashima,
  • Yoshihiko Saito

DOI
https://doi.org/10.1016/j.ijcha.2015.06.003
Journal volume & issue
Vol. 8, no. C
pp. 122 – 127

Abstract

Read online

Objective: In optical coherence tomography (OCT)-guided percutaneous coronary intervention (PCI), stent size is usually determined according to the pre-PCI lumen size of either the distal or proximal reference site. However, the effect of the OCT imaging catheter crossing the target lesion on the reference lumen measurements has not been studied. We evaluated changes in the reference lumen size before and after PCI using frequency domain OCT. Methods: For 100 consecutive patients with PCI, mean lumen diameter (LD) and lumen area (LA) were measured at the proximal and distal reference sites before and after coronary stent implantation with OCT. Results: Mean LD and LA of the distal reference site were significantly increased after PCI with stent implantation (2.57 ± 0.6 to 2.62 ± 0.64 mm, p 1.50 mm2. Tissue characteristics were not correlated with changes in reference lumen size. Conclusions: When we select the stent size during OCT-guided PCI, we need to pay attention to the decrease in the luminal measurement of the reference sites, especially in lesions with tight stenosis.