PLoS ONE (Jan 2019)

Clinical evaluation of General Electric new Swiftscan solution in bone scintigraphy on NaI-camera: A head to head comparison with Siemens Symbia.

  • F Thibault,
  • M Bailly,
  • G Le Rouzic,
  • G Metrard

DOI
https://doi.org/10.1371/journal.pone.0222490
Journal volume & issue
Vol. 14, no. 9
p. e0222490

Abstract

Read online

PurposeThe General Electric (GE) Swiftscan solution combines a new Low Energy High Resolution and Sensitivity collimator (LEHRS) with image processing (Clarity 2D) and tomographic step and shoot continuous mode. The aim of this study was to compare clinical and physical performances of this new technology in bone scintigraphy.MethodsPhysical phantom measurements were performed using GE LEHRS, GE Low Energy High Resolution (LEHR) and Siemens LEHR collimators. These measurements were associated with a prospective clinical study. Sixty-seven patients referred for bone scintigraphy were enrolled from February to July 2018. Each patient underwent two acquisitions consecutively on GE and Siemens gamma camera, using respectively Swiftscan solution and LEHR collimator.ResultsOn planar acquisitions, maximum sensitivity was 100 cts/MBq for Siemens LEHR. GE SwiftScan LEHRS and GE LEHR maximum sensitivity were respectively 9% and 22% lower. Using Clarity 2D, GE Swiftscan LEHRS spatial resolution was the best with 9.2 mm versus 10.1 mm and 10.6 mm for GE LEHR and Siemens LEHR collimators. In tomographic mode, the sensitivity of GE Swiftscan solution was superior to both LEHR systems (16% and 25% respectively for Siemens and GE). There was no significant difference in spatial resolution. In clinical use, signal was higher on Siemens system and noise was lower on GE Swiftscan solution. Contrast-to-noise ratios were not significantly different between the two systems. There was a significant image quality improvement with GE SwiftScan in planar images and in whole body scan. No significant difference in image quality was observed on SPECT images.ConclusionNew GE SwiftScan collimator design improved sensitivity compared to "classical" GE LEHR collimator without compromising resolution. GE SwiftScan solution enhances planar image quality with a better Clarity 2D resolution recovery and noise treatment. In SPECT mode, GE SwiftScan solution improves volumetric sensitivity without significant impact on image quality, and could lead to time or dose reduction.