Química Nova (Oct 2003)
Aplicação da CLAE para determinação do ácido 10-Hidróxi-2-decenóico (10-HDA) em geléia real pura e adicionada a mel brasileiro HPLC aplication for 10-HDA determination in pure royal jelly and honey with royal jelly
Abstract
The aim of this paper was to determine the 10-HDA in pure royal jelly and products containing royal jelly, using HPLC methodology. 10-HDA is the natural indicator of the presence of royal jelly in products and also gives the authenticity of pure royal jelly. The chromatographic conditions used were: isocratic system, C18-H column, auto sampler, diode array UV-VIS detector (225 nm), mobile phase with methanol/water (45:55), pH= 2.5 and a-naphtol as internal standard. The results obtained using laboratory samples for pure royal jelly were 2.37%, varying from 0.15% for honey with 10% of royal jelly to 2.10% for honey with 90% of royal jelly respectivelly. For commercial products, the 10-HDA content varied from no detectable to 0.026%. The recovery test presented a minumum of 100.44% The detection limit was 45.92 ng/mL and the quantification limit was 76.53 ng/mL.