Read online

In this work, a double-cylinder viscometer is designed to measure dynamic viscosity over a pressure range from atmospheric pressure up to 150 MPa and a temperature range of 278.15–333.15 K. A high-pressure closed cavity is designed innovatively and the magnetic coupling is adopted to transfer the torque to reduce the friction; the inner cylinder with ruby bearing is designed to reduce the friction torque, thus the accuracy of the viscosity measurement is improved. The experiment of measuring the standard viscosity liquid (N10 and N35) under normal pressure and measuring the viscosity of methylbenzene under the pressure of 0.1–150 MPa were carried out, and considering all the experimental data, the uncertainty of the viscosity measurements is approximately ±3%.