Radiation (Oct 2022)
In Vivo Dosimetry for Superficial High Dose Rate Brachytherapy with Optically Stimulated Luminescence Dosimeters: A Comparison Study with Metal-Oxide-Semiconductor Field-Effect Transistors
Abstract
The purpose of the study was to calibrate and commission optically-stimulated luminescence dosimeters (OSLDs) for in vivo measurements in contact-based 192Ir treatments for superficial high dose rate (HDR) brachytherapy in place of metal-oxide-semiconductor field-effect transistors (MOSFETs). Dose linearity and dose rate dependence were tested by varying source-to-OSLD distance and dwell time. Angular dependence was measured using a solid water phantom setup for OSLD rotation. A group of OSLDs were readout 34 consecutive times to test readout depletion while OSLDs were optically annealed using a mercury lamp for 34.7 h. End-to-end tests were performed using a Freiburg flap and Valencia applicator. OSLD measurements were compared to MOSFETs and treatment planning system (TPS) doses. OSLD response was supralinear for doses above 275 cGy. They were found to be independent of dose rate and dependent on the incident angle in edge-on scenarios. OSLDs exhibited minimal readout depletion and were successfully annealed after 24 h of illumination. Freiburg flap measurements agreed well with the TPS. For the Valencia, OSLDs showed to be the more accurate system over MOSFETs, with a maximum disagreement with the TPS being 0.09%. As such, OSLDs can successfully be used in place of MOSFETs for in vivo dosimetry for superficial HDR brachytherapy.