CVIR Endovascular (Dec 2020)

Temperature-induced configuration changes in hydrogel-coated coils and their relevance in embolization procedures

  • Ruben Lopez Benitez,
  • Tomas Reyes del Castillo,
  • Levent Kara,
  • Joachim Kettenbach,
  • Justus Roos

DOI
https://doi.org/10.1186/s42155-020-00189-0
Journal volume & issue
Vol. 3, no. 1
pp. 1 – 7

Abstract

Read online

Abstract Background The present study attempted to demonstrate how the configuration of hydrogel-coated coils is influenced by different temperature exposures. Thirty detachable hydrogel-coated coils were evaluated in an in vitro water immersion test under five different temperature ranges (22.6 °C, 37 °C, 40–50 °C, 50–60 °C, and 60–70 °C). The configuration changes were classified (configuration I, configuration II, and configuration III) according to the curling that occurred during 30 min of immersion. Configuration stability of five Hydrogel-coated coils was also evaluated in a two-step temperature immersion test. Results All hydrogel-coated coils showed some configuration changes during water immersion. However, a logarithmic transformation of the time and temperature data showed a significant (p < 0.05) negative linear correlation between time and temperature for all coil configurations (configuration I: R = 0.97, configuration II: R = 0.98, configuration III: R = 0.97). The time needed to reach configuration III (complete coiling) was 160.4 ± 41.9 s at 37.5 °C (range: 100–205 s), 45.7 ± 22.2 s at 47.5 °C (range: 23–70 s), 20.2 ± 7.2 s at 57.5 °C (range: 14–32 s), and 10.3 ± 2.4 s at 67.5 °C (range: 7–13 s). Conclusions Temperatures above 55 °C induced immediate configurational changes in the hydro-coated coils, achieving complete curling within less than 30 s. Temperatures near 36 °C (normal body temperature) require more time to reach optimal coil curling (configuration III). The optimization of HydroCoil preparation can reduce interventional procedural time and improve clinical results.

Keywords