Engineering Proceedings (Oct 2023)

Compact Thermographic Device with Built-in Active Reference Element for Increased Measurement Accuracy

  • Michal Švantner,
  • Vladislav Lang,
  • Jiří Skála,
  • Tomáš Kohlschütter,
  • Jan Šroub,
  • Lukáš Muzika,
  • Jan Klepáček,
  • Milan Honner

DOI
https://doi.org/10.3390/engproc2023051017
Journal volume & issue
Vol. 51, no. 1
p. 17

Abstract

Read online

Many common thermographic cameras have quite good sensitivity (0.05 °C or better) but limited accuracy, often about ±2 °C. This is not sufficient for quantitative measurements such as human body temperature diagnostics, where an accuracy of 0.3 °C is assumed. A thermographic device with a built-in active reference element for enhanced precision measurement was developed for these purposes. It was adapted for human body temperature diagnostics using face temperature measurements. It is based on a micro-bolometers detector and the reference element is heated to a temperature of 37 °C, which is mostly assumed as decisive for an increased temperature indication. As the reference element is embedded into the device housing, the use of the device can be very flexible. The performed human temperature measurement experiments showed that the accuracy of the introduced device is comparable with the thermographic measurements obtained using an external black body, which is often used for these applications.

Keywords