BMJ Open Ophthalmology (Oct 2020)

Validating e-norms methodology in ophthalmic biometry

  • H John Shammas,
  • Joe F Jabre

DOI
https://doi.org/10.1136/bmjophth-2020-000500
Journal volume & issue
Vol. 5, no. 1

Abstract

Read online

Objective To validate the extrapolated norms or e-norms methodology in establishing a reference range for the biometric data used for intraocular lens power calculation.Methods and Analysis All measurements were performed with an optical low-coherence reflectometer. A novel technique, the e-norms methodology, was used to determine the normative values of measurements.Results Eyes (n=500) were measured to evaluate the axial length (AL), K readings (Ks), anterior chamber depth (ACD) and lens thickness (LT). Using the e-norms methodology, the normal AL ranged from 22.50 to 24.50 mm (mean=23.50 mm), with medium-long eyes between 24.51 and 24.99 mm and the long eyes measuring 25.00 mm and longer; the medium-short eyes ranged from 22.01 and 22.49 mm, with the short eyes measuring 22.00 mm and shorter. Normal values ranged from 2.50 to 3.50 mm for ACD (mean=3.00 mm), from 4.40 to 5.44 mm for LT (mean=4.92 mm), and from 42.50 to 44.82 dioptres for Ks (mean=43.66 dioptres).Conclusion Measurements of the biometric mean values compared favourably with published data. The e-norms methodology assisted in establishing a biometric reference range. Furthermore, it allowed us to cluster patients into groups based on AL differences.