Iranian Journal of Medical Physics (Oct 2018)

Comparison of Image Quality of Low Voltage 64-slice Multidetector CT Angiography with the Standard Condition in Patients Suspected of Pulmonary Embolism

  • Mohammad Davoodi,
  • Soudabeh Belash-Abadi,
  • Seyed Hossein Aalaei-Andabili,
  • Ali Kabir

DOI
https://doi.org/10.22038/ijmp.2018.27829.1302
Journal volume & issue
Vol. 15, no. 4
pp. 237 – 242

Abstract

Read online

Introduction: Reduction of peak kilovoltage (kV) setting has been a useful approach to d creating radiation dose; however, it may have varied effects on noise and the accuracy of diagnosis. Thus, we compared image quality between low (80 kV) and standard kilovoltage (100 kV) protocols. Material and Methods: This triple blind non-randomized parallel quasi-experimental study was conducted on 140 cases of questionable pulmonary embolism. Results: Image quality was twice as high as the standard protocol in the 80-kV group (odds ratio=2.08). Main, segmental, and subsegmental arteries showed significantly higher vascular enhancement (P