Annals of Hepatology (Jan 2025)
Diagnostic performance of FibroTouch® in assessing hepatic steatosis and fibrosis in patients with metabolic dysfunction-associated steatotic liver disease: An Asian experience
Abstract
Introduction and Objectives: FibroTouch® has shown efficacy in staging hepatic fibrosis in patients with chronic viral hepatitis B, but its performance in assessing liver steatosis and fibrosis in metabolic dysfunction-associated steatotic liver disease (MASLD) patients remains understudied. We aimed to evaluate the diagnostic performance of FibroTouch® in assessing steatosis and fibrosis in the MASLD population. Materials and Methods: Liver stiffness measurements and steatosis were assessed using FibroTouch® and FibroScan®, with FibroScan® as the reference standard. Pearson's correlation test evaluated correlations, and kappa statistics determined agreement between the two methods. Optimal cut-off values of FibroTouch® for predicting hepatic steatosis and fibrosis stages were determined through ROC curve analysis with the Youden index method. Results: Strong correlations were observed between FibroTouch® UAP and FibroScan® CAP (rho=0.74) and LSM values (rho=0.87) (p < 0.001 for both) in a total of 380 patients. The mean CAP value for the entire cohort was 285 ± 51 dB/m, and the median LSM for the cohort was 5 .3kPa. The optimal FibroTouch® UAP cutoffs were 229 dB/m for S0 vs. S1, 267 dB/m for S1 vs. S2, and 294 dB/m for S2 vs. S3. For FibroTouch® LSM, the optimal cutoffs were 6.0 kPa for F0-F1 vs. F2, 7.9 kPa for F2 vs. F3, and 10.6 kPa for F3 vs. F4. Moreover, FibroTouch® effectively assessed hepatic steatosis and fibrosis in patients with different BMIs. Conclusions: FibroTouch® proved valuable in assessing hepatic steatosis and liver fibrosis staging in MASLD patients, enhancing its applicability in various clinical settings as a suitable and convenient option for MASLD patients.
Keywords