Abstract

Read online

In recent years, MALDI-TOF Mass Spectrometry (MS) method has emerged as a promising and a reliable tool for bacteria identification. In this study we compared Bruker MALDI-TOF MS and conventional phenotypic methods to identify a collection of 333 Gram-positive clinical isolates comprising 22 genera and 60 species. 16S rRNA sequencing was the reference molecular technique, and rpoB gene sequecing was used as a secondary gene target when 16Sr RNA did not allow species identification of Corynebacterium spp. We also investigate if score cut-offs values of ≥ 1,5 and ≥ 1,7 were accurate for genus and species-level identification using the Bruker system. Identification at species level was obtained for 92,49% of Gram-positive rods by MALDI-TOF MS compared to 85,89% by phenotypic method. Our data validates the score ≥ 1,5 for genus level and ≥ 1,7 for species-level identification in a large and diverse collection of Gram-positive rods. The present study has proved the accuracy of MALDI-TOF MS as an identification method in Gram-positive rods compared to currently used methods in routine laboratories.