The Scientific World Journal (Jan 2015)

Comparison of the Proseal, Supreme, and I-Gel SAD in Gynecological Laparoscopic Surgeries

  • Sanli Mukadder,
  • Begec Zekine,
  • Kayhan Gulay Erdogan,
  • Ozgul Ulku,
  • Ucar Muharrem,
  • Yologlu Saim,
  • Durmus Mahmut

DOI
https://doi.org/10.1155/2015/634320
Journal volume & issue
Vol. 2015

Abstract

Read online

We compared proseal, supreme, and i-gel supraglottic airway devices in terms of oropharyngeal leak pressures and airway morbidities in gynecological laparoscopic surgeries. One hundred and five patients undergoing elective surgery were subjected to general anesthesia after which they were randomly distributed into three groups. Although the oropharyngeal leak pressure was lower in the i-gel group initially (mean ± standard deviation; 23.9 ± 2.4, 24.9 ± 2.9, and 20.9 ± 3.5, resp.), it was higher than the proseal group and supreme group at 30 min of surgery after the trendelenburg position (25.0 ± 2.3, 25.0 ± 1.9, and 28.3 ± 2.3, resp.) and at the 60 min of surgery (24.2 ± 2.1, 24.8 ± 2.2, and 29.5 ± 1.1, resp.). The time to apply the supraglottic airway devices was shorter in the i-gel group (12.2 (1.2), 12.9 (1.0), and 6.7 (1.2), resp., P=0.001). There was no difference between the groups in terms of their fiber optic imaging levels. pH was measured at the anterior and posterior surfaces of the pharyngeal region after the supraglottic airway devices were removed; the lowest pH values were 5 in all groups. We concluded that initial oropharyngeal leak pressures obtained by i-gel were lower than proseal and supreme, but increased oropharyngeal leak pressures over time, ease of placement, and lower airway morbidity are favorable for i-gel.