Hecheng xiangjiao gongye (Dec 2024)

"Application of halogen moisture meter in volatile content measurement of nitrile rubber"

  • XU Bin, YUAN Ji-yao, XIAO Ye, CHEN Dong-ping, YING Chan-juan, GUAN Yan

DOI
https://doi.org/10.19908/j.cnki.ISSN1000-1255.2023.06.0504
Journal volume & issue
Vol. 46, no. 6
pp. 505 – 508

Abstract

Read online

The volatile content in nitrile rubber was measured by halogen moisture meter and compared with the results obtained by the hot-roll method. The influence of different measurement conditions on the measured results of the halogen moisture meter was investigated. The results showed that using the halogen moisture meter, the volatile content in nitrile rubber raw rubber could be measured by both the preset drying time method and preset mass loss method. Under suitable conditions such as sample mass of 3 g, drying temperature of 110 ℃, preset heating time of 8 min, or preset mass loss of 1/140 mg/s, the measured results were consistent with the hot-roll method. This method reduced measure time, was easy to operate, and could be used to guide quickly actual production.