Journal of Food and Pharmaceutical Sciences (Jan 2014)
Analytical Method Validation of Selected Organochlorine Pesticides in Fortified Rice Using Gas Chromatography with Electron Capture Detector
Abstract
Gas chromatography with electron capture detector (GC-ECD) has been validated for simultaneous determination of selected organochlorine pesticides namely heptachlor, endosulfan, dieldrin, dan p,p’-dichlorodiphenyltrichloroethane (p,p’-DDT) using aldrin as standard internal. Some parameters for analytical method validation were evaluated, i.e. linearity, precision, sensitivity expressed with limit of detection and limit of quantification, and recovery. GC-ECD was linear for determination of these pesticides with r values ≥ 0.99. The values of LOD obtained were 2.50, 2.97, 1.61, and 2.54 ng/mL for heptachlor, endosulphan, dieldrin, and p,p’-DDT, respectively. Furthermore, the relative standard deviation (RSD) values obtained for the assessment of analytical precision were lower than those required by Horwitz. Finally, the used GC-ECD method was accurate for determination of these organochlorine psticides in fortified rice samples which were indicated by the acceptable recovery percentages.