Chinese Journal of Magnetic Resonance (Dec 2017)

A Magnetic Resonance Imaging Receiver Design Based on NI PXIe-7966R

  • HU Jin-jie,
  • YAO Jun-jiang,
  • XU Jun-Cheng,
  • LIU Jun,
  • WANG Zi,
  • JIANG Yu

DOI
https://doi.org/10.11938/cjmr20172565
Journal volume & issue
Vol. 34, no. 4
pp. 489 – 497

Abstract

Read online

A magnetic resonance imaging receiver design based on NI PXIe-7966R is proposed, with which the magnetic resonance signals are sampled directly and down-converted digitally, the raw data are uploaded and the magnetic resonance image are restored. The system-level digital signal processing (DSP) development tools offered by NI LabVIEW field programmable gate array (FPGA) was used for FPGA function modeling, simulation and automatic code generation of hardware description language (HDL). It was very flexible during the digital down conversion (DDC) designing. The sampling rate of this module was 50 Mbps, and the receiver bandwidth could be varied between 100 Hz and 1 MHz. The experimental results showed that the receiver design is a high performance magnetic resonance receiver solution.

Keywords