BioTechniques (Jun 2013)

Continuous Fluorescence Monitoring of Rapid Cycle DNA Amplification

  • Carl T. Wittwer,
  • Mark G. Herrmann,
  • Alan A. Moss,
  • Randy P. Rasmussen

DOI
https://doi.org/10.2144/000114043
Journal volume & issue
Vol. 54, no. 6
pp. 314 – 320

Abstract

Read online

Rapid cycle DNA amplification was continuously monitored by three different fluorescence techniques. Fluorescence was monitored by (i) the double-strand-specific dye SYBR Green I, (ii) a decrease in fluorescein quenching by rhodamine after exonuclease cleavage of a dual-labeled hydrolysis probe and (iii) resonance energy transfer of fluorescein to Cy5 by adjacent hybridization probes. Fluorescence data acquired once per cycle provides rapid absolute quantification of initial template copy number. The sensitivity of SYBR Green I detection is limited by nonspecific product formation. Use of a single exonuclease hydrolysis probe or two adjacent hybridization probes offers increasing levels of specificity. In contrast to fluorescence measurement once per cycle, continuous monitoring throughout each cycle monitors the temperature dependence of fluorescence. The cumulative, irreversible signal of hydrolysis probes can be distinguished easily from the temperature-dependent, reversible signal of hybridization probes. By using SYBR Green I, product denaturation, annealing and extension can be followed within each cycle. Substantial product-to-product annealing occurs during later amplification cycles, suggesting that product annealing is a major cause of the plateau effect. Continuous within-cycle monitoring allows rapid optimization of amplification conditions and should be particularly useful in developing new, standardized clinical assays.