Sensors (Jan 2022)

Experimental Assessment of Cuff Pressures on the Walls of a Trachea-Like Model Using Force Sensing Resistors: Insights for Patient Management in Intensive Care Unit Settings

  • Antonino Crivello,
  • Mario Milazzo,
  • Davide La Rosa,
  • Giacomo Fiacchini,
  • Serena Danti,
  • Fabio Guarracino,
  • Stefano Berrettini,
  • Luca Bruschini

DOI
https://doi.org/10.3390/s22020697
Journal volume & issue
Vol. 22, no. 2
p. 697

Abstract

Read online

The COVID-19 outbreak has increased the incidence of tracheal lesions in patients who underwent invasive mechanical ventilation. We measured the pressure exerted by the cuff on the walls of a test bench mimicking the laryngotracheal tract. The test bench was designed to acquire the pressure exerted by endotracheal tube cuffs inflated inside an artificial model of a human trachea. The experimental protocol consisted of measuring pressure values before and after applying a maneuver on two types of endotracheal tubes placed in two mock-ups resembling two different sized tracheal tracts. Increasing pressure values were used to inflate the cuff and the pressures were recorded in two different body positions. The recorded pressure increased proportionally to the input pressure. Moreover, the pressure values measured when using the non-armored (NA) tube were usually higher than those recorded when using the armored (A) tube. A periodic check of the cuff pressure upon changing the body position and/or when performing maneuvers on the tube appears to be necessary to prevent a pressure increase on the tracheal wall. In addition, in our model, the cuff of the A tube gave a more stable output pressure on the tracheal wall than that of the NA tube.