International Journal of Implant Dentistry (Sep 2020)

Implant primary stability depending on protocol and insertion mode — an ex vivo study

  • Henning Staedt,
  • Peer W. Kämmerer,
  • Elisabeth Goetze,
  • Daniel G. E. Thiem,
  • Bilal Al-Nawas,
  • Diana Heimes

DOI
https://doi.org/10.1186/s40729-020-00245-3
Journal volume & issue
Vol. 6, no. 1
pp. 1 – 9

Abstract

Read online

Abstract Background Dental implant primary stability is thought to be a fundamental prerequisite for the long-term survival and success. The aim of this study was to analyze the influence of protocol and insertion mode on dental implant stability ex vivo. One hundred and twenty implants were inserted either manually or machine-driven into porcine mandibles by a standard or over-dimensioned protocol. Dental implant stability was measured via resonance frequency analysis (RFA), insertion torque (IT), and torque out (TO). Results Statistically significant higher IT and TO values were seen after standard protocol insertion (p < 0.05), whereas manual and machine-driven insertion mode showed equivalent values. Conclusions The over-dimensioned protocol exceeded the primary stability values recommended for immediate implant insertion; therefore, it could be recommended as well.