Applied Sciences (Feb 2021)

UHPLC-MS Method for the Analysis of the Molecular Adjuvant Sulfavant A

  • Genoveffa Nuzzo,
  • Emiliano Manzo,
  • Marcello Ziaco,
  • Laura Fioretto,
  • Ana Margarida Campos,
  • Carmela Gallo,
  • Giuliana d’Ippolito,
  • Angelo Fontana

DOI
https://doi.org/10.3390/app11041451
Journal volume & issue
Vol. 11, no. 4
p. 1451

Abstract

Read online

A fast and sensitive method that is based on Ultra High Performance Liquid Chromatography coupled with High Resolution Mass Spectrometry (UHPLC-HRMS) for the measurement of Sulfavant A, a molecular adjuvant with a sulfolipid skeleton, is described. The method has been validated over the linearity range of 2.5–2000 ngmL−1 using a deuterated derivative (d70-Sulfavant A) as internal standard. Chromatographic separation is based on a UHPLC Kinetex® 2.6 µm PS C18 column and a gradient of methanol in 0.32 mM ammonium hydroxide solution buffered at pH 8. The lowest limit of quantification of Sulfavant A was 6.5 ngmL−1. The analytical procedure was tested on an extract of mice lung spiked with 30, 300, and 1500 ng of Sulfavant A. The analysis revealed a precision and accuracy value (as a mean value of all the quality control samples analyzed) of 4.7% and 96% in MeOH and 6.4% and 93.4% in the lung extracts, respectively.