Micromachines (Feb 2023)

Hemisphere Tabulation Method: An Ingenious Approach for Pose Evaluation of Instruments Using the Electromagnetic-Based Stereo Imaging Method

  • Zhongjie Long,
  • Yongting Chi,
  • Dejin Yang,
  • Zhouxiang Jiang,
  • Long Bai

DOI
https://doi.org/10.3390/mi14020446
Journal volume & issue
Vol. 14, no. 2
p. 446

Abstract

Read online

Drilling of a bone surface often occurs in clinical orthopaedic surgery. The position and orientation of the instrument are the most important factors in this process. Theoretically, some mechanical components may assist in orienting an instrument to certain bone shapes, such as the knee joint and caput femoris. However, the mechanical assisting component does not seem to work in some confined spaces where the bone shape is a free-form surface. In this paper, we propose an ingenious hemisphere tabulation method (HTM) for assessing the pose accuracy of an instrument. The acquisition and assessment of HTM is conducted based on an electromagnetic-based stereo imaging method using a custom-made optical measurement unit, and the operation steps of HTM are described in detail. Experimental results based on 50 tests show that the HTM can identify ideal poses and the evaluated pose of an instrument location on a hemisphere model. The mean error of pose localisation is 7.24 deg, with a range of 1.35 to 15.84 and a standard of 3.66 deg, which is more accurate than our previous method.

Keywords