Current Directions in Biomedical Engineering (Sep 2016)

A method to determine the kink resistance of stents and stent delivery systems according to international standards

  • Brandt-Wunderlich Christoph,
  • Schwerdt Christopher,
  • Behrens Peter,
  • Grabow Niels,
  • Schmitz Klaus-Peter,
  • Schmidt Wolfram

DOI
https://doi.org/10.1515/cdbme-2016-0064
Journal volume & issue
Vol. 2, no. 1
pp. 289 – 292

Abstract

Read online

The kink behavior of vascular stents is of particular interest for clinicians, stent manufacturers and regulatory as a kinked stent generates a lumen loss in the stented vessel and can lead to in-stent restenosis. In this study methods to determine the kink resistance of stents and stent delivery systems according to the ISO 25539-2 and FDA guidance no. 1545 were presented. The methods are applicable for balloon expandable stents as well as for self-expanding stents and determine the lumen loss and residual diameter change dependent on the specific bending radius.

Keywords